PHILIPS Super Rotalix Metal (SRM) rotating anode X-ray tube
Tube RX à anode tournante
Drehanodenröhre in Glas/Metallbauweise
This tube uses a metal housing for the area around the anode and cathode. The X-rays
leave the tube via a beryllium window added to the metal housing. The metal part of the tube is
earthed and thus captures stray electrons; this makes it possible to bring the tube collimator much
closer to the anode than in conventional tubes, which results in a reduction in the effects of off-focal radiation.
As there is no danger of breakdown of the tube due to the vaporization (as in the case of a glass
insert), it is possible to heat the filament to higher temperatures. This leads to an improvement in rating.
The Super Rotalix Metal tube was introduced in 1973.
Hauteur • Height • Höhe : 34 cm • 1' - 1" 4/10
Diagram of Philips Rotalix metal tube and casing