VICTOREEN 1B85 THYRODE Thin Wall Beta-Gamma Counter Tube
				  Compteur Geiger-Müller sensible aux rayonnements Bêta et Gamma
				  Beta-Gamma-Zählrohr
              Direct substitute replacement for LND 725 and similar GM detectors used in GM probes such as Ludlum 44-38 and Ludlum 44-6, Eberline HP-270, etc.
			  • Operating voltage: 850 ... 950 V
              • Min beta energy: 160 keV
              • Max threshold voltage: 800 V
              • Min plateau length: 200 V
              • Plateau slope: 3%/100 V
              • Recovery time: 100 μs
              • Background (at 900 V): 40 counts/minute
              • Life (at 6000 counts per minute): 108 counts
              • Wall: aluminum, 30 mg/cm2
              • Gas: argon plus organic quench
              Fiche technique • Data sheet • Datenblatt:
              1B85.pdf
               
			  
              Longueur • Length • Länge : 10,6 cm • 4" 1/8
              Diamètre • Diameter • Durchmesser : 19 mm • 3/4"
			  
			  
              
 