This tube shows a typical failure mode of X-Ray tubes: the anode track is damaged by overheating and roughened up. If the quantity of heat delivered during an individual exposure exceeds the track capacity, the anode surface can melt, as shown below.

 

General Electric CRT 1-2 Rotating Anode X-Ray Tube

Longueur • Length • Länge : 35 cm • 13" 3/4
Diamètre • Diameter • Durchmesser : 124 mm • 2" 7/8

General Electric CRT 1-2 Rotating Anode X-Ray Tube

General Electric CRT 1-2 Rotating Anode X-Ray Tube

General Electric CRT 1-2 Rotating Anode X-Ray Tube

General Electric CRT 1-2 Rotating Anode X-Ray Tube

General Electric CRT 1-2 Rotating Anode X-Ray Tube

RADIOLOGY, Volume 37, Number 2, August 1941

 